INSERT PURGE CASSETTE
1.
Open the purge cassette package.
2.
If included, discard the purge extension line and secure the yellow luer
connector on the purge tubing.
3.
Spike the glucose bag/bottle.
4.
Press the NEXT soft button to continue.
5.
Open the purge cassette door by pressing the release on the left side of
the controller.
6.
Insert the purge cassette into the compartment on the front of the
controller. Follow the diagram on the inside of the purge cassette door
for proper placement.
7.
Slide the purge disc into the slot to the right of the purge cassette
until it snaps into place. The controller will automatically begin
priming the purge cassette.
8.
Extend the purge tubing and close the purge cassette door. There is
sufficient room around the edges of the purge cassette door so that it
will not pinch the purge tubing as it exits.
9.
The controller automatically begins priming the purge cassette after it
is inserted. The progress bar marks the progress of the purge cassette
priming.
CONNECT THE IMPELLA CATHETER
1.
Remove the Impella Catheter from its package using sterile technique
and inspect the catheter for damage.
2.
Inspect the cable for damage, including damage to the connector pins at
the controller end.
3.
Pass the sterile connector cable from the Impella Catheter off the sterile
field.
4.
Open the cover on the blue catheter plug by rotating clockwise.
Line up the notch on the connector cable with the notch in the blue
catheter plug on the front of the Automated Impella Controller and
plug the cable into the controller.
Impella 5.5
®
with SmartAssist
®
Circulatory Support System
5.
Once the purge cassette is primed and the controller detects that the
connector cable is plugged in, it prompts you to connect the luer to
the Impella Catheter.
6.
If included, disconnect the Y-connector and purge extension line from
the purge tubing and discard. Connect and tighten the yellow luer on
the purge tubing to the Impella Catheter purge arm if not already
complete.
7.
When the controller detects that the luer is connected, it automatically
begins priming the purge lumen.
8.
Once the purge lumen is primed, the controller automatically advances
to the next screen.
9.
The first step on the next screen prompts you to enter the purge
fluid information.