Rise time (T10...90)
1) 2)
CO
2
Range
Accuracy
Rise time (T10...90)
1) 2)
N
O
2
Range
Accuracy
Rise time (T10...90)
1) 2)
Anesthetic agents
Range
Halothane
Isoflurane
Enflurane
Sevoflurane
Desflurane
Accuracy
Rise time (T10...90)
Automatic detection
Primary gas
Secondary gas
The mentioned identification thresholds (primary and secondary) apply to rising agent concentrations (e.g., beginning of an anes-
thetic case).
When anesthetic agent concentration decreases (e.g., end of an anesthetic case), the gas analyzer measures agent concentra-
tions down to 0,05 Vol% based on the last identified agent. Below that value the gas analyzer indicates loss of agent identification
and a value of 0 Vol%.
Respiratory rate (RR)
Range
Accuracy
Resolution
1) The accuracy of the measurement of O
expiratory values specified in the technical data apply at a maximum respiratory rate of 60 /min with an I:E ratio of 1:1. The influence of respiratory rate
and I:E ratio on the accuracy has been determined for the measured gas concentrations in a simulated breathing system with step changes in concentra-
tion.
2) The specified accuracy is obtained at the latest 450 seconds after start-up.
EMC declaration for IEC 60601-1-2, Edition 3.2
General information
The device was tested for electromagnetic compatibility using
accessories from the list of accessories. Other accessories may
only be used if they do not compromise the electromagnetic com-
patibility. The use of non-compliant accessories may result in
increased electromagnetic emissions or decreased electromag-
netic immunity of the device.
Electromagnetic environment
The medical device is intended for use in an electromagnetic envi-
ronment as specified below. The user must ensure its use in such
an environment.
Supplement Scio Four modules
<500 ms
0 to 10 Vol%
±(0.43 Vol% + 8 % rel.)
<300 ms
0 to 100 Vol%
±(2 Vol% + 8 % rel.)
<300 ms
0 to 8.5 Vol%
0 to 8.5 Vol%
0 to 10 Vol%
0 to 10 Vol%
0 to 20 Vol%
±(0.20 Vol% + 15 % rel.)
<450 ms
At the latest at 0.3 Vol%
At the latest at 0.4 Vol%
With a Desflurane concentration greater than 4 Vol%, mixture detection occurs at
the latest when the concentration of the second anesthetic agent rises above 10 %
of the Desflurane concentration.
0 to 100 /min (Respiratory rate is derived from CO
0 to 60 /min: ±1 /min with an I:E ratio of 1:1
>60 /min: not specified
0 to 80 /min: ±1 /min with an I:E ratio of 1:2
>80 /min: not specified
1 /min
, CO
, N
O and anesthetic agents depends on the respiratory rate and on the I:E ratio. The accuracies of the
2
2
2
The device may be used in the direct vicinity of other devices only
if Dräger has authorized this device arrangement. If no authoriza-
tion has been given by Dräger, it must be ensured that the device
functions correctly in the desired arrangement before using it. The
instructions for use for the other devices must be followed.
English
real-time data)
2
15

Advertenties