EN
GTIN
Batch number
Serial number
Production date
Catalogue number
Customer number
Compliance with
(EU) 2017/745
Compliance with
2010/35 EU
Note: The content of the marking is always the same, the position on the valve may vary
depending on the type.
A – INLET STEM
The combination valve is fi tted to the gas cylinder by a threaded inlet stem. The inlet stem can
be taper threaded or parallel threaded with diff erent sizes depending on the cylinder size and
material.
B – FILLING PORT
A fi lling port is provided for fi lling the gas cylinder at a fi lling station, it has no function to patient
use. The fi lling port contains a non-return valve (NRV). The NRV prevents fl ow of gas without
the use of specifi c fi lling adaptors which enables venting/evacuating gas from the cylinder and
fi lling gas into the cylinder.
C – PRESSURE INDICATOR
The pressure indicator displays the amount of gas in the gas cylinder. The pressure indicator is
of an active type which means it always indicates amount of gas in the gas cylinder whether the
shut-off valve is opened or closed.
D – SHUT-OFF VALVE
The combination valve is provided with a shut-off valve to isolate the gas in the cylinder from
the rest of the combination valve functions. It must be opened during cylinder fi lling and gas
supply. Part of the handwheel is an indicator showing open/close status of the shut-off valve.
MARKING DETAIL
Compliance with CDG TPE
(Amendment) & (EU Exit) Regulations
2020
4/293
Inlet connection type
Filling port type
Total weight allowance
of the package
UDI code
Inlet pressure
Compliance with
valve standards