5. MEDICAL DEVICE DESCRIPTION
A medical combination valve combines the function of shut-off valve of a high-pressure gas
cylinder and pressure regulator for the use with medical gases. Gas from the cylinder is first
controlled by the shut-off valve and then passed through the pressure regulator to the flow
outlet or the pressure outlet, either of which can supply gas to a patient. The outlet pressure is
fixed and downstream overpressure protected by a low-pressure relief valve.
There are three basic alternatives:
•
combination valve with a quick coupler outlet, the outlet pressure is fixed and the flow
provided is on the demand of the downstream equipment, (there is no selectable flow outlet)
•
combination valve with flow outlet, a selectable range of flows are controlled by the
combination valve and can be changed by the flow control head, (there is no quick coupler
pressure outlet)
•
the combination valve containing both quick coupler outlet and flow outlet.
Detailed information about the performance and technical data related to your device can
be found in Appendix Nr. 1. Appendix Nr. 1 is in form of symbols. Explanations for the symbols
are available in Chapter 12 - Glossary.
SHUT OFF VALVE STATUS INDICATOR:
fully OPEN
green colour
fully CLOSED
red colour
partly OPEN
red / green colour
H
E
D
Typical MediVital® valve configuration
3/293
EN
F
G
I
B
C
A
J