4. Product description
Head End Cover. They contain a low-voltage,
safe 24V transformer. The electric motors/
actuators, the battery backup, the ACK and
handsets (24V) are connected to the Control
Box and Junction Boxes via dust and moisture
proof plugs and cords/cables.
5. Battery Backup allows for off-mains
emergency operation for up to 15 minutes
under a normal load. To ensure long battery
life, the bed should be connected to mains
power, at all possible times. Do not exceed
the duty cycle.
6. Electrical protection of wiring and cords will
differ between Class 1 (Earthed) and Class 2
(Double Insulation) Beds.

4.3 Standard features

The backrest is a large mattress panel that raises
from a supine position to an upright position,
convenient for sitting, which allows a patient
to enjoy the flexibility of changing to multiple
positions for comfort and health.
Kneebreak
The kneebreak is a double mattress panel
that splits as it electronically lifts or bends
the patient's thigh and calf at the knee, thus
combining functionality and comfort for both
the patient and carer.
Auto-Contour
A Handset feature that uses one button to adjust
both Backrest & Kneebreak simultaneously, into
a cardiac chair position, even when the bed is in
it's lowest height position.
Battery Backup
The Battery Backup allows for operation of the
bed, should mains power fail, or if the bed is
being operated temporarily away from mains
power. The battery can operate for a total
time of up to 15 minutes under a normal load.
Make sure not to exceed the duty cycle (2 min
operation, 18 minutes rest).
www.humancaregroup.com
To ensure long battery life, the bed should be
connected to mains power, at all possible times.
If the backup battery becomes flat, raising
lowering functions will slow down or cease. A
replacement should be ordered and fitted.
Batteries in good condition will normally
recharge in 12 hours or overnight when the bed
is plugged back into the mains power supply.
Extended periods without mains power will
make the batteries unable to be recharged or
operate at all. A replacement unit will need to
be ordered and fitted.
Backlit Handset
The Backlit Handset permits patients to control
and adjust their own bed, to different positions,
to suit their changing comfort levels throughout
their stay in bed. The soft lighting feature allows
a patient the convenience of night-time use,
without other lights.
Cardiopulmonary Resuscitation (CPR) Quick
Release System (only for the FloorLine-i Plus bed)
The CPR emergency procedure (combined
rescue breathing and chest compressions) is
used on a patient in cardiac arrest. Firstly, using
either one of the CPR Quick Release Levers,
located under either side of the backrest, the
backrest can be quickly flattened, mechanically.
Secondly, the CPR button on the ACK, will also
electrically flatten the bed into the defined
(country specific) CPR position. The bed ordered
will be country-appropriate!