9. Operation Instructions
horizontal/flat position, then secondly flattens
out the Backrest & Kneebreak simultaneously.
Single Button Trendelenburg - Press the
Single-Button-Trendelenburg for an immediate
Trendelenburg position (Head Down/Feet High).
9.8 Mattress Platform Positions
• True Floor-Level - 99mm (4") off floor
• Highest Platform Height 799mm (31½")
• Backrest Panel - 70º
• Thigh Panel - 45º }operate together as
• Calf Panel- 25º }the kneebreak panel
• Kneebreak Position 110º
• Trendelenburg Tilting Position 18º
• Reverse-Trendelenburg Tilt Position 18º
• Head High Trendelenburg Tilt Position 45º
• Head Low Trendelenburg Tilt Position 45º
9.9 Digital control box functions
• Battery Back-Up System
• CPR Quick Release System (only P5)
• Patient Egress/Under Bed Light (Only P5)
• Service Monitor (option)
9.10 Castors and brakes
The FloorLine-i Plus and FloorLine-i bed is fitted
with Dual Wheel Castors as standard.
The set of standard wheel castors comprises:
• 1 black Steering/Directional Castor
• 3 grey Braked Castors
26
Directional lock
The Directional Lock (black) is located on one
castor at the head end. It can be locked in any
position to aid and ensure a smooth and straight
path down a hallway.
The Directional Lock Castor should be engaged
to the "ON" position at all times. To disengage
the Directional Lock, depress the "OFF" lever.
Braked castors
Braked Castors (grey) are locked by depressing
the front of the pedal on each castor lever.
This is done by foot pressure when wearing
appropriate protective shoes.
Brakes are released by depressing the top of the
castor lever until it unlocks.
Do not attempt to set or release
o
the brakes using your hands or
fingers as injury may occur.
9.11 Moving the bed
Before moving the bed, raise
o
the mattress platform to a
minimum of 200mm from the
floor.
• When moving the bed, first release the
brakes on all castors so that the bed can
move freely.
• If brakes are not released and the bed is
forcibly moved when the wheels are locked,
the bed may be damaged.
• Re-engage the directional lock to assist with
steering.
• To control movement forwards or backwards
only, orientate the directional/steering
wheel, so it is parallel with the side edge of
the bed and then depress the lock lever to
the "ON" position with your foot. This will
engage a lock on this wheel which will permit
the bed to run straight forward or backward
ENG