9. Operation Instructions
CPR Quick Release Levers
Pull either one of the RED CPR Quick Release
Levers, located under either side of the backrest.
This will immediately (mechanically/manually)
lower the Backrest panel to a flat position.
When using the Emergency Release
Levers, DO NOT stop the backrest panel
until it has fully lowered, or actuator
damage may occur.
CPR button on the ACK
The bed will quickly and electrically level the
bed into the horizontally flat CPR position.
The sequence of movements firstly levels the
mattress platform to a horizontal/flat position,
then secondly flattens out the Backrest &
Kneebreak simultaneously. (Note: Early models
may have a different sequence!)
Regularly check the mechanics (see below)
of the CPR Quick Release Levers to ensure
the system is always operating correctly.
If intubation or head access is required...
1. Raise the bed, if it is at Floor-Level.
2. Remove the Headboard. There is no need
to place the headboard under the mattress,
as all Human Care bed platforms are solid
enough for effective chest compressions.
3. Release the Brakes, to allow the bed to be
pulled out from the wall for easy access to the
patient's head. Once the bed is positioned,
reapply the brakes.
CPR Emergency Quick Release System, Testing
Operation & Making Adjustments
1. Raise the back rest using the handset, while
applying slight hand force to the top of the
back rest panel. The back rest should raise to
it's highest position; if it doesn't then the CPR
cable is too tight and the release mechanism
is partially engaged. The cable needs to be
loosened via an adjustment (Photo 1).
www.humancaregroup.com
Photo 1.
2. Check the CPR cable tension, it should not
be too tight, you should be able to move the
cable laterally slightly (Photo 2).
Photo 2.
3. The CPR cable is adjusted by loosening the
two locking nuts and moving the threaded
section in the required direction, 'In' to loosen
or 'Out' to tighten (Photo 3). Access to both
locking nuts can be gained from underneath
the mattress platform.
Photo 3.
4. Repeat Step 1, the back rest should raise to
it's highest position with a slight downward
hand force applied.
5. To ensure that the quick release system is
operating properly, raise the back rest to it's
highest position with the handset then by
using slight downward hand force on the top
of the back rest panel with one hand lift the
red CPR lever upward with the other using the
inside of your fingers (Photo 4). The back rest
panel should lower quickly.
ENG
29