9. Operation Instructions
when being moved. The wheel itself can be
positioned towards the front or the back of
the castor outrigger but it is recommended
to point down the bed's length.
For safety, if leaving the bed
o
for any time ensure castors are
turned inwards and locked.
9.12 Folding the bed - (Requires two
persons)
The FloorLine-i Plus and FloorLine-i beds are
designed to be folded for transport, relocation
and storage.
Requirements
1. Two qualified assembly persons
2. A mains power outlet/power-point
Tools required:
1. 3mm & 5mm Allen Keys
2. 13mm & 17mm spanner/socket wrench
3. Packing tape for securing panels
4. Small flat screwdriver
Folding stages
1) Preparation
2) Disconnect Cables
3) Loosen Bed Base Bolts
4) Remove Mattress Platform from Bed Base
5) Fold Mattress Platform
6) Fold Bed Base
7) Ready bed for storage
Preparation
1. Remove all bedding and accessories. (e.g.
Mattress, Side Rails IV/Self Help Poles).
2. Remove the Head/Footboards and store.
3. Using the handset, test all functions of the
bed. Complete a visual inspection. Report any
faults or damage.
www.humancaregroup.com
4. Clean the bed thoroughly.
5. Ensure there is sufficient floor space for
disassembly.
6. Disassemble on a covered surface (e.g.
a blanket), to protect both floor and bed
surfaces.
7. Connect the mains power cable to the mains
power outlet.
8. Ensure all castors are locked for safety.
9. Using the handset, flatten/neutralize all
the bed positions (backrest, kneebreak,
Trendelenburg/reverse).
Disconnect cables
1. To allow better access to cables and bolts:
1) Raise the bed to its maximum height and
2) Lift backrest and foot-end panels
2. Release the backrest and kneebreak cables
from the plastic retaining clips (1 backrest
cable clip/3 kneebreak cable clips) that secure
them to inside of the mattress platform
frame.
3. Disconnect the kneebreak (long) cable from
its mini-fit actuator plug, by first removing
the plastic retaining clip with a small flat
screwdriver. Keep the plastic retaining clips
safe by storing them inside the mini-fit plugs.
Keep the cable safe, during the disassembly
process.
4. Repeat above step for backrest (short) cable.
Note: Nothing needs to be done with any
other cables, as they all remain connected
to the control box and attached to the bed
base during storage.
Loosen the bed base bolts
1. To aid bed base folding later, at each pedestal
end, use the 17mm spanner, to completely
loosen all 8 Bed Base Lower Beam Bolts (4 x
Hinging Bolts and 4 x Pedestal Latch Locking
Bolts), that are seated in either end's Pedestal
End Lower Brackets. DO NOT REMOVE THE
NUTS from the end of these Bolts.
ENG
27