8. Assembly/Disassembly
4. Lift backrest and foot-end panels, allowing
easy upper beam access.
5. At the foot end, lift the beams to slot the 2
Platform Bracket's Fixing-Bolts (Nylon Bush)
into the 2 Pedestal End's Fixed J-Hooks.
Head End Swinging J-Hook
(Growth Link Plate)
Foot End Fixed J-Hook
Platform Nylon
Bush Fixing-Bolts
Lower Beam - 4
Nyloc Nuts
6. Lift the beams to fully seat the fixing-bolts.
7. Position the nylon bush to the outside, and the
thin spacer/washer between the metal brackets.
Spacer variations exist on early models.
8. Repeat above steps at the head end.
9. Using the 5mm Allen Key, firmly tighten all 4
Upper Beam Fixing-Bolts, ensuring that each
bush is correctly located in its J-Hook.
10. Using the 13mm spanner, confirm that all 4
Fixing-Bolt Nuts are firmly attached to the end
of the bolt, with a wide space between the nut
and the upper beam bracket.Note: The Head
End Swinging J-Hooks must be swung up, then
down into the nylon bushes. Also, the head end
fixing-bolts have a wider spacer.
DO NOT TIGHTEN these Fixing-Bolt Nuts! The Swinging J-Hook
(Growth Link Plate) and the additional space between the
nuts and the brackets allow for platform flexing and full bed
operation particularly for Trendelenburg positions.
Connecting cables (FloorLine-i)
Cables connect the Control Box with:
1. The Mains Power Outlet
2. The Handset
3. The 4 Actuators (Hi-Lo Head, Hi-Lo Foot,
Backrest, Kneebreak).
20
• All cables are pre-attached to the Control
Box, which is located on the inside of the
Head End Pedestal on the Bed Base.
• The two Hi-Lo Actuators are located inside
the Head and Foot End Pedestals and cables
are pre-attached to their actuators.
• The Backrest & Kneebreak Actuators are
located on the underside of the mattress
platform. Their respective cables must be
attached to their appropriate mini-fit plugs.
1. Attach the mains cable to the mains power
outlet, rechecking for damage in transit.
2. To allow better access to the underneath of
the mattress platform:
• Use the handset to raise the bed to its
maximum height
• Lift backrest and foot-end panels
3. Connect the Backrest (short) cable to the
short mini-fit plug cable, and lock in the plastic
retaining clips, ensuring that the locking tabs sit
in the plug's slots securely, as pictured.
4. Connect the Kneebreak (long) cable to
the long mini-fit plug cable and lock in the
retaining clips, as pictured.
5. Secure all cables to the inside frame of the
mattress platform with the plastic 'p' clips.
Connecting cables (FloorLine-i Plus)
Cables run between the Control Box and the
various actuators. The cables and Backrest &
Kneebreak actuators are attached to the upper
and lower beams of the bed frame.
1. If used, detach the additional Junction Box
and re-attach the Backlit Handset Control.
ENG
www.humancaregroup.com