Guidance and Manufacturer's Declaration – Emissions All Equipment and Systems
Rapid Infuser, RI-2 is intended for use in the electromagnetic environment specified below. The
®
The Belmont
customer or user of the Belmont
Emissions Test
Compliance
RF Emissions
Group 1, Class A
CISPR 11
Harmonics IEC
Complies or Not
61000-3-2
applicable
Flicker
Complies or Not
IEC 61000-3-3
applicable
Guidance and Manufacturer's Declaration—Immunity All Equipment and Systems
Rapid Infuser, RI-2 is intended for use in the electromagnetic environment specified below. The
®
The Belmont
customer or user of the Belmont
Immunity Test
IEC 60601 Test Level
Electrostatic
±6kV contact
Discharge (ESD)
IEC 61000-4-2
±8kV air
Radiated RF
3 V/m
IEC 61000-4-3
Electrical Fast
±2kV on AC Mains
Transient/burst
IEC 61000-4-4
Surge
±1kV Differential
IEC 61000-4-5
±2kV Common
Conducted RF
3 Vrms from
IEC 61000-4-6
150 kHz to 80 MHz
Power Frequency
3A/m
50/60Hz Magnetic
Field
IEC 61000-4-8
Voltage dips, short
>95% Dip for 0.5 Cycle
interruptions and
voltage variations
60% Dip for 5 Cycles
on power supply
input lines
30% Dip for 25 Cycles
IEC 61000-4-11
>95% Dip for5 Seconds
Hoofdstuk 4: Parameterinstellingen En Preventief Onderhoud
Table 201
Rapid Infuser, RI-2 should assure that it is used in such an environment.
®
Electromagnetic Enforcement – guidance
Rapid Infuser, RI-2 uses RF energy only for its internal
®
The Belmont
function. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
Complies
Complies
Table 202
Rapid Infuser, RI-2 should assure that it is used in such an environment.
®
Compliance Level
±6kV contact
±8kV air
3 V/m
±2kV on AC Mains
±1kV Differential
±2kV Common
V1 = 3 Vrms
3A/m
>95% Dip for 0.5 Cycle
60% Dip for 5 Cycles
30% Dip for 25 Cycles
>95% Dip for 5 Seconds
43
Electromagnetic Environment -
Guidance
Floors should be wood, concrete or
ceramic tile. If floors are synthetic, the
relative humidity should be at least
30%.
Radiated RF frequency is negligible at
any single frequency
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment.
If interference occurs, it may be
necessary to position the Belmont
Rapid Infuser, RI-2 further from
sources of power frequency magnetic
field.
Power frequency magnetic fields
should be that of a typical location in a
typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the
Rapid Infuser, RI-2
®
Belmont
requires continued operation during
power mains interruptions, it is
recommended that the Belmont
Rapid Infuser, RI-2 be powered
from an uninterruptible power supply
or battery.
®
®