Status Flags are generated at the end of a calibration attempt and can be used to help determine the cause of a failed calibration. The Status Flag
will appear on Results screens when calibration fails and will appear near the bottom of a Genius™ Checker/Calibrator Test Report. The Status
Flags are listed in the following table:
Status Flag
001
003
00A
012
01A
022
042
062
082
102
182
202
400
800
Section XI - Customer Service
The Genius™ Checker/Calibrator requires factory calibration and inspection once a year. Notify and return the device to Covidien on or before the
calibration date for the Genius™ Checker/Calibrator. The Genius™ Checker/Calibrator will display a message one month prior to the calibration
date as a reminder.
Caution: There are no serviceable parts inside, return to factory for service.
In the event that it is necessary to return a unit for repair, please observe the following:
1. Contact Covidien as shown below, or your local customer service representative for an Authorized Return Number and shipping instructions.
United States
1-800-964-5227
Parts Listing
Please contact your local customer service center or sales representative for the parts listed below.
Description
Genius™ Checker/Calibrator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .303097
Refurbished Genius™ Checker/Calibrator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .303096
Section XII - Warranty
Covidien warrants the Genius™ Checker/Calibrator to be free of defects in materials and workmanship under normal use and service for a period
of 1 year from the date of delivery by Covidien to the first purchaser. Liability hereunder is limited to repair or replacement, at Covidien's option,
at Covidien service facility for any product, which shall under normal use and service, appear to Covidien to have been defective in material or
workmanship. This warranty shall not apply to, and Covidien shall not be responsible for, any loss arising in connection with the purchase of any
product that has been repaired by anyone other than Covidien or its authorized representative, or which has been subject to misuse, neglect,
or accident, or which has been used otherwise than in accordance with the instructions furnished by Covidien. Covidien neither assumes nor
authorizes any representative or other person to assume for it any liability other than as expressly set forth herein.
Covidien disclaims all other warranties, expressed or implied, including any implied warranty of merchantability or fitness for a
particular purpose or application other than as expressly set forth in the product labeling. In no event will Covidien be liable for any
incidental, indirect, or consequential damages in connection with the purchase or use of this product.
Genius
Checker/Calibrator
TM
Status Flag Description
Passed – No recalibration
Calibration Passed
Calibration Failed - Fine accuracy limit exceeded (avg) on low target
Calibration Failed - Fine accuracy limit exceeded (avg) on high target
Calibration Failed - Fine accuracy limit exceeded (avg) on low target and fine accuracy limit exceeded (avg) on high target
Calibration Failed - Span error of thermometer readings detected on low target
Calibration Failed - Span error of thermometer readings detected on high target
Calibration Failed - Span error of thermometer readings detected on low target and high target
Calibration Failed - Slope out of range
Calibration Failed - Offset out of range
Calibration Failed - Slope out of range and offset out of range
Calibration Failed - Post calibration variances are not within range
Calibration Failed - Thermometer reading on Low target is not within 1.4 deg C (2.5 deg F) accuracy limit
Calibration Failed - Thermometer reading on High target is not within 1.4 deg C (2.5 deg F) accuracy limit