Download Inhoudsopgave Inhoud Print deze pagina

Safe Working Load (Swl) - Howard Wright M9 TRANSFER Gebruiksaanwijzing

Inhoudsopgave

Advertenties

Beschikbare talen

Beschikbare talen

1.4 APPLICATION ENVIRONMENT
The stretcher is only intended for use in these environments:
APPLICATION ENVIRONMENT 1:
Intensive/critical care provided in a hospital where 24 hour medical supervision and constant
monitoring is required and provision of life support system/equipment used in medical procedures is
essential to maintain/improve the vital functions of the patient.
APPLICATION ENVIRONMENT 2:
Acute care provided in a hospital or other medical facility where medical supervision and monitoring
is required and medical equipment used in medical procedures may be provided to help maintain or
improve the condition of the patient.
APPLICATION ENVIRONMENT 5
Outpatient (ambulatory) care, which is provided in a hospital or other medical facility, under medical
supervision where ME EQUIPMENT, is provided for the need of persons with illness, injury or
disability for treatment, diagnosis or monitoring
NOTE: This includes use in nursing homes, rehabilitation and geriatric facilities.

1.5 SAFE WORKING LOAD (SWL)

Maximum patient weight = SWL minus weight of accessories
1 x O2 cylinder
1 x IV pole
1 x PH pole
1 x Mattress
MAXIMUM PATIENT WEIGHT
MAXIMUM SWL
220kg
MAX
Figure 1. Safe Working Load
NOTE: This is an indicative list. Other accessories can be used and the weight of these must be deducted.
=
12 kg
=
1.2 kg
=
6.1 kg
=
12 kg
30 kg
=
SWL - 30 kg
=
220 kg
250kg