7. ACCESSORIES
7.11 RESIDUAL CURRENT DEVICE (IF FITTED)
WARNING: If pressing the TEST button and the indicator light does
not change from red to black, do not use the stretcher
and consult either a Biomedical engineer or an approved
electrician.
A Residual Current Device (RCD) is used to increase the electrical safety
of an electrical appliance.
The RCD should be tested each time the stretcher is plugged in to the
mains supply.
Testing the RCD:
1. Connect the power cord to the mains supply and
The indicator light should illuminate red.
2. Press the TEST button - indicator light should change from red to black.
3. Press the RESET button to resume operation.
NOTE:
Please see Residual Current Device Instructions for use
(M799-81) for further information.

7.12 PLUG IN EXTENSION

The plug in extension (M921-03) can be fitted at the foot end of the stretcher.
The extension is for supporting the feet of taller patients.
To fit:
1. Hold the extension with both hands.
2. Lower the legs into the sockets on the deck.
2. Place the bolster mattress on the extension.
To remove:
1. Remove the bolster mattress.
2.
Remove the extension from the sockets
26
press the RESET button.
M9 Transfer Instructions For Use (M999-190, Version 1, 2022-04)
TEST
RESET
INDICATION LIGHT
Figure 38. Residual Current Device (RCD)
Figure 39. Plug in Stretcher extension