e) Bed end
Check that the bed end can be removed and reinstalled. If jamming occurs clean the bed end sockets and
bed end pegs and add a small amount of Vaseline (petroleum jelly) or food grade grease to the sockets.
f) Castors
Set the castor pedal in the brake position. Check that all four castors cannot swivel or roll.
Set the castor pedal in the neutral position. Check that all four castors are free to swivel and roll.
Set the castor pedal in the neutral position. Check that there is no excessive play in the fifth wheel mechanism
and the 5th wheel itself.
Set the castor pedal in the steer position. Check that the 5th wheel (if fitted) locks in line with the length
of the stretcher and cannot swivel. Check that the remaining castors can swivel.
Set the castor pedal in the steer position. Check that the steering castor (if fitted) locks in line with the
length of the stretcher and cannot swivel. Check that the remaining castors can swivel.
g) Mechanical items
Check that all bolts, screws, nuts and pivot-pins are securely fastened.
Check both siderails. Raise siderail and check gas strut assists raising. Check siderail does not lower
unassisted. Check that the siderail latch automatically latches with the siderail in the raised position – this can
be seen visually and heard audibly. Lift the siderail latch and lower the siderail to see that it is in good working
order.
h) Accessories
Check the siderail covers for rips, cracks or other damage.
Check the mattress (and bolster mattress) covers for rips, cracks or other damage.
Check that the IV pole is straight and otherwise free of damage. Check that the locking collar securely locks
the pole when tightened.
Check that the PH pole is straight and otherwise free of damage.
Check that the PH pole strap is free from cuts and abrasions.
Check that the PH handle strap adjustment mechanism is operating correctly (applies to adjustable PH handle
option).
Check the blanket cradle for damage.
Check the oxygen cylinder carrier for damage.
Check the drainage bag hook rail for damage.
Check the urine bottle carrier for damage.
Check the monitor tray for damage. Check that the strap is free from cuts and abrasions and that the buckle is
in working order.
Check the over bed table for damage. Check that the height of the table can be adjusted.
Check the orthopaedic frame for damage.