To lower:
1. Use one hand to lift the release lever (where it says "LIFT TO LOWER SIDERAIL") latch.
2. Use the other hand to push the top rail of the siderail towards the foot end and down into the collapsed position.
NOTE:
Use bed ends with siderails to improve patient safety and reduce the risk of patient falls.
Lift to lower siderail
Figure 20. Lift to lower siderail label
Lift to lower siderail
Figure 21. Siderail in full height
6.11 UNIVERSAL ACCESSORY MOUNT RAIL
WARNING: The safe working load of the Universal Accessory Mount Rail is 25kg.
WARNING: Maximum eccentric load application to the Universal Accessory Mount Rail is 100mm.
WARNING: Ensure accessories are fitted securely to the Universal Accessory Mount Rail as per manufacturers
instructions.
The Universal Accessory Mount Rail provides a secure rail system for standard accessories to be fitted.
The Universal Accessory Mount Rail is located at the head end of the stretcher.
NOTE:
Maximum torque to be applied to the Universal Accessory Mount Rail is 24 Nm
Figure 22. Universal Accessory Mount Rail
1. Lift release lever
M9 Transfer Instructions For Use (M999-190, Version 1, 2022-04)
6. OPERATION
2. Push down top rail.
Accessory
Mount Rail
19