9.3 CLEANING-DISINFECTION-STORAGE OF THE HANDPIECE AND ELECTRICAL CABLE

Steps
What to do
1
Cleaning
To clean the screw core of the motor and
metallic parts, use the disposable brush
(#19149) moistened with cleaning solutions.
Dip the blade oscillation shaft of the motor
in a sterile distilled water bath and run the
motor back and forth 10 times.
Remove the motor from the cup, unplug it
from the console, and wipe the oscillation
shaft of the motor with a lint-free cloth.
Cable can be cleant with a lint-free cloth
moistened with cleaning solutions.
Air-dry the shaft of the motor with clean,
filtered compressed air (medical compressed
air).
Disinfecting & Drying
2
• Wipe each part of the motor with a lint-free
cloth moistened with disinfecting solutions.
• Use appropriate disinfecting solutions (spray
or wipe clean disinfectant) according to the
manufacturers' instructions.
• Dry carefully with disposable instrument
cleaner (lint-free) then with clean, filtered,
compressed air.
3
Storage
Document Aletiq : Rév. J-05.2023 (Validée)
10
MORIA
#65039-J-05.2023
IMPORTANT WARNINGS
The handpiece must be carefully cleaned
after each surgical procedure.
Do not use abrasives or scrapers to clean
any elements of the system. Doing so may
result in reduced precision, burrs and/or
irregular tissue resection.
Do not immerse the motor in the sterile
distilled water bath.
To prevent damaging the connectors and
to ensure proper functioning of the motor,
never pull on the cables and never hold the
motor by its cable.
It is imperative to verify that the devices are
perfectly clean and free of any organic or
other residues.
Do not gas sterilize (ETO) the motor.
Do not autoclave the motor.
No current sterilisation process is
compatible with MORIA motors.
Failure to completely dry the inside of the
motor can result in oxidation.
Do not store products that are not
completely dry; doing so may create rust
and result in irregular corneal resection.
When not in use, the One Use-Plus system
should be kept in its storage box #22519514
in a dry atmosphere.