7 OPERATION

Steps
What to do
1
Preliminary check before any usage
2
Connect the motor to the control
unit, select One Use-Plus and the
speed, then check vacuum prior use
3
Insert the head in the rails of the
suction ring
4
Check for keratome translation prior
to use (forward and backward pass)
5
Position the suction ring on the eye.
Activate vacuum by pressing the
"Vacuum" footswitch once.
6
Check the intraocular pressure with
the tonometer
7
Lubricate the suction ring, the
head and the blade
8
Once the One Use-Plus system is
in position, activate the motor by
pressing the "Forward" footswitch.
When the microkeratome head
touches the stop, immediately
release the "Forward" footswitch.
Reverse by pressing the "Backward"
footswitch.
9
Release the "Vacuum" footswitch by
pressing once.
Remove the suction ring.
Document Aletiq : Rév. J-05.2023 (Validée)
8
MORIA
#65039-J-05.2023
IMPORTANT WARNINGS
It is imperative to verify that the devices are perfectly
clean and free of any organic or other residues.
Also, please refer to EVOLUTION 3 or 3E user
manual (#65038, 65051, 65060) for operations
prior use. The operator should be familiar with the
functioning of the EVOLUTION 3 or 3E control unit
(#19360, 19380).
If you have additional questions, consult the MORIA
website or your MORIA distributor.
See EVOLUTION 3 or 3E console user manual
(#65038, 65051, 65060).
Refer to the indicative nomogram for the selection of
speed.
When inserting the head in the rails of the suction
ring, the handpiece must be at a 45° angle from the
threaded shaft of the ring, to avoid any blade damage
If the bottom of the head (the area where the blade
edge protrudes) is in contact with the threaded
shaft of the suction ring, inadvertent damage to the
blade may occur. Release vacuum and change to a
completely new set to avoid the possibility of irregular
cut.
Once the head properly inserted, slightly reduce the
angle from 45° to 0°
Carefully engage the head into the ring
The shaft of the ring must be aligned with the motor.
Press vacuum ON, make forward then backward
movements using footpedals, then press vacuum
OFF. Please refer to EVOLUTION 3 or 3E user
manual (#65038, 65051, 65060)
Do not use the One Use-Plus system if the forward
pass is not smooth, regular and uninterrupted.
Verify that the device will not interfere with any other
medical devices used during the procedure.
The tonometer (#19042) must be perfectly dry and
used only on dry eyes. If the pressure is below 65 mm
Hg, do not proceed with the surgery.
Prior to use, lubricate the rails of suction ring,
head, and blade with balanced salt solution or
another appropriate solution.
The device should not be placed in contact with
solutions for which the potential interactions are
unknown.
See EVOLUTION 3 or 3E console user manual
(#65038, 65051, 65060).
Hold the suction ring by its handle; check that your
fingers do not hinder the forward movement of the
device.
Check nothing prevents or modifies head movement
in the rails of the suction ring
Ensure that there are no obstructions (speculum,
eyelashes, eyelids, etc.) in its path.
Discard the disposable head, ring and aspiration
tubing after each procedure in an appropriate
container.