Section 2
Installation
®
Brooks
VDM300 Vapor Delivery Module

2-7 Diagnostic Port Connection

2-8 Return Shipment

2-9 Transit Precautions

2-10 Removal from Storage

2-8
All manuals and user guides at all-guides.com
The VDM300 can also be operated in a Service Mode through the micro
USB diagnostic port connection utilizing Brooks Expert Support Tool
(BEST) and a USB to micro USB cable. You should not use this port for
normal operation. The BEST software installed on a PC acts as the
interface to perform these operations. For more information on the use of
the BEST, see Section 4-5 Operation of the Brooks Expert Support Tool
(BEST) for VDM300 later in this manual and/or the Brooks Expert Support
Tool (BEST) manual available with the purchase of the BEST.
Prior to returning any device to the factory, visit the Brooks web site
(www.BrooksInstrument.com) for a Return Materials Authorization Number
(RMA#), or contact one of the locations provided on p. 2-1
Prior to returning the device, it must be drained & purged in accordance
with the following:
All flow devices returned to Brooks require completion of the Brooks
Instrument Decontamination Statement, along with a Safety Data Sheet
(SDS) for the fluid(s) used in the instrument. Failure to provide this
information will delay processing by Brooks personnel.
Copies of these forms can be downloaded from the Brooks website
(www.brooksinstrument.com/globalsupportcenters) or are available from
any of the Brooks Instrument locations provided on the back cover.
To safeguard against damage during transit, transport the device to the
installation site in the same container used for transportation from the
factory, if circumstances permit.
Upon removal of the device from storage, a visual inspection should be
conducted to verify its "as-received" condition. If the device has been
subject to storage conditions in excess of those recommended (refer to "2-
3 Recommended Storage Practice" on p. 16), it should be subjected to a
pneumatic pressure test in accordance with applicable vessel codes.
Installation and Operation Manual
X-VAP-VDM300-eng
Part Number: 541B210AAG
August, 2017