Section 2
Installation
®
Brooks
VDM300 Vapor Delivery Module
2-3 Recommended Storage Practice
2-4 VDM300 Mounting and Orientation
2-5 Mechanical Connections
2-2
All manuals and user guides at all-guides.com
If intermediate or long-term storage of the device is required, it is
recommended that it be stored in accordance with the following conditions:
• Within the original shipping container.
• Ambient temperature 21°C (70°F) nominal, 32°C (90°F) maximum, 7°C
(45°F) minimum.
• Relative humidity 45% nominal, 60% maximum, 25% minimum.
Figure 2-1 shows the proper mounting orientation for the VDM300.The
VDM300 should be mounted using four 5/16" (7.9mm) screws tightening
them to the screw manufacturer's recommended specifications.
Figure 1-6 shows the mechanical dimensions of the VDM300
The recommended installation procedure guidelines are as follows:
• The device should be located in a clean, dry atmosphere relatively free
from shock and vibration.
• Install the device in such a manner that permits easy purge and removal if
the device requires servicing.
Prior to installation, ensure that all piping is clean and free from
obstructions. Install piping in such a manner that permits easy access to
the device if removal becomes necessary.
Installation and Operation Manual
X-VAP-VDM300-eng
Part Number: 541B210AAG
August, 2017