This service information describes the connection of the IntelliBridge to the Dräger Evita 4 ventilator.
Data communication
The IntelliBridge EC10 module supports RS232 serial data communication with the following
specifications for connection with the external device:
• Baud rate:
• Word length:
N O T E
RS232 parameters must be set and verified according to the Dräger Evita 4 ventilator service manual
by authorized personnel.
Instruction for connection
Connect the IntelliBridge EC5 D-Sub 9-pin male connector to one of the three RS232 connectors of
the Dräger Evita 4 ventilator as shown in the photo below. Make sure to tighten the screws of the
connector properly.
IntelliBridge EC5 connection to RS232 ports
1
2
Service Information
7
19200
8 bits
1
1
1
Dräger Evita 4 ventilator - RS232 (COM) port
IntelliBridge EC5 ID module
453564161101
• Stop bits:
• Parity:
2
7
1
Even
27