Service Information; Data Communication; Instruction For Connection - Philips IntelliBridge ED203 Gebruiksaanwijzing

Apparaatdriver voor de dräger evita 4-ventilator
Inhoudsopgave

Advertenties

This service information describes the connection of the IntelliBridge to the Dräger Evita 4 ventilator.

Data communication

The IntelliBridge EC10 module supports RS232 serial data communication with the following
specifications for connection with the external device:
• Baud rate:
• Word length:
N O T E
RS232 parameters must be set and verified according to the Dräger Evita 4 ventilator service manual
by authorized personnel.

Instruction for connection

Connect the IntelliBridge EC5 D-Sub 9-pin male connector to one of the three RS232 connectors of
the Dräger Evita 4 ventilator as shown in the photo below. Make sure to tighten the screws of the
connector properly.
IntelliBridge EC5 connection to RS232 ports
1
2

Service Information

7
19200
8 bits
1
1
1
Dräger Evita 4 ventilator - RS232 (COM) port
IntelliBridge EC5 ID module
453564161101
• Stop bits:
• Parity:
2
7
1
Even