Storage
• Do not store in damp conditions (such as in a
bathroom) and do not transport with any damp
items.
• When storing and transporting, protect from
prolonged direct sunlight.
• The parts of the atomiser set can be securely
stored in the storage compartment
the device in a dry place, ideally in the original
packaging.
10. ATOMISER AND RE-
PLACEMENT PARTS
The atomiser set parts and replacement parts are
available from the corresponding service address
(according to the service address list). Please state
the corresponding order number.
Designation
Atomiser set = year pack
includes:
Mouthpiece
Nosepiece
Silicone adult mask
Silicone children's mask
Atomiser
Compressed air hose
Filter
Angled fitting
Nasal douche
If the device is not used according to the
specifications, perfect functionality cannot
be guaranteed! We reserve the right to make
technical changes to improve and develop the
product. This device and its atomiser set con-
form with the European standard EN 60601-
1-2 (Group 1, class B, in accordance with
CISPR 11, IEC 61000-3-2, IEC 61000-3-3,
IEC 61000-4-2, IEC 61000-4-3, IEC 61000-4-
4, IEC 61000-4-5, IEC 61000-4-6, IEC 61000-
4-7, IEC 61000-4-8, IEC 61000-4-11) and is
subject to particular precautions with regard
to electromagnetic compatibility. Please note
that portable and mobile HF communication
systems may interfere with this device. For
more details, please contact our Customer
Services at the address indicated.
. Store
4
Material
REF
603.76
PP/silicone
PP/silicone
Silicone/PP
Silicone/PP
PP/silicone
PVC
PU
PP
PP
601.37
Notes on electromagnetic
compatibility
• The device is suitable for use in all environ-
ments listed in these instructions for use, in-
cluding domestic environments.
• The use of the device may be limited in the
presence of electromagnetic disturbances. As
a result, the device may fail.
• Avoid using this device directly next to other
devices or stacked on top of other devices, as
this could lead to faulty operation. If, however,
it is necessary to use the device in the manner
stated, this device as well as the other devices
must be monitored to ensure they are working
properly.
• The use of an atomiser set other than that
specified or provided by the manufacturer of
this device can lead to an increase in electro-
magnetic emissions or a decrease in the de-
vice's electromagnetic immunity; this can re-
sult in faulty operation.
• Keep portable RF communication devices (in-
cluding peripheral equipment, such as anten-
na cables or external antennas) at least 30 cm
away from all device parts, including all cables
included in delivery.
• Failure to comply with the above could impair
the performance of the device.
11. WHAT IF THERE ARE
PROBLEMS?
Problems/
Possible cause/remedy
questions
The atomiser
1. Too much or too little medica-
produces no
or too little
aerosol.
2. Check nozzle for blockages.
IMPORTANT: Carefully pierce
the fine holes only from the
underside of the nozzle.
3. Atomiser not held vertically.
4. Unsuitable medication fluid
The medication fluid should be
prescribed by a doctor.
24
tion in the atomiser.
Minimum: 2 ml,
Maximum: 8 ml.
Clean nozzle if necessary (e.g.
by rinsing out). Then start
using the atomiser again.
added for nebulisation (e.g.
too viscous).