7.4 Inserting the valve
• Press the valve
11
for this purpose on the upper side of the atom-
iser
.
9
Only use the valve
. Never use it with the masks
13
never with the nosepiece
Connecting the atomiser set to the
atomiser
• Connect the atomiser
iser set (mouthpiece
, silicone children's mask
14
).
16
The most e ective form of nebulisation is by
using the mouthpiece. Nebulisation using a
mask is only recommended if it is not pos-
sible to use a mouthpiece (e.g. for children
who are not yet able to inhale using a mouth-
piece).
When using a mask to inhale, take care to
ensure the mask fits well and the eyes are
unobstructed.
• Before the treatment, pull the atomiser up-
wards out of the holder
• Start the nebuliser using the On/O control
• Spray mist pouring out indicates that the de-
vice is operating correctly.
7.5 Treatment
• When inhaling, sit upright and relaxed at a table
and not in an armchair to avoid compressing
the airways and therefore impairing the e ec-
tiveness of the treatment.
• Breathe in the atomised medication deeply.
Important
The device is not suitable for continuous operation;
after 30 minutes of operation it must be switched
o for 30 minutes.
During the treatment, hold the atomiser
straight (vertically), otherwise the atomisation
will not work and faultless functionality is not
guaranteed.
Important
Essential oils, cough syrups, gargling solutions and
drops to be used as a rub or in a steam bath are
wholly unsuitable for inhalation using a nebuliser.
These additives are often viscous and can impair
the correct functioning of the device and therefore
the e ectiveness of the application in the long
into the opening intended
with the mouthpiece
11
14 15
.
16
to the desired atom-
9
, silicone adult mask
13
or nosepiece
15
.
7
term. For individuals with a hypersensitive bron-
chial system, medications containing essential
oils may under certain conditions cause an acute
bronchospasm (a sudden cramp-like restriction of
the bronchi with shortness of breath). Consult your
doctor or pharmacist in relation to this matter.
and
7.6 Stopping the inhalation
Once the mist is only coming out in an irregular
flow or if the sound changes when inhaling, you
can stop the treatment.
• Switch o the nebuliser after treatment using
the On/O control
the mains.
• Place the atomiser
ter the treatment.
7.7 Nasal douche
You can use the nasal douche to clean your nasal
cavities. The nasal douche creates a fine jet of liq-
uid for cleaning and moistening the nasal mucous
membranes.
Warning
• Do not use the nasal douche in combination
with other medication or essential oils.
.
3
Important
• Only use the nasal douche with isotonic saline
solutions.
• Replace the nasal douche after a year.
The maximum volume of the nasal douche is 10 ml.
If you are unsure about the correct dilution of the
saline solution, please consult your pharmacist.
7.8 Using the nasal douche
To use the nasal douche correctly, proceed as fol-
lows
:
B
1. Twist the cover
move it in an upward direction.
2. Also lift the collection container
3. Fill the saline container
4. Place the collection container
container
22
5. Place the cover
Twist the cover
it clicks firmly into place.
6. Connect the compressed air hose
line container
7. Carefully position the cover
nostrils. Breathe through your nose.
21
and disconnect it from
3
back in its holder
9
90° anticlockwise and re-
20
21
with saline solution.
22
into the saline
21
.
on the saline container
20
in a clockwise direction until
20
from below.
22
in one of your
20
af-
7
out.
.
22
to the sa-
8