Cleaning
Important
The device must be switched o , disconnected
from the mains and allowed to cool down each
time before cleaning.
The atomiser
and the rest of the atomiser
9
set used, such as the mouthpiece
, atomiser
, nasal douche
15
9
cleaned with hot but not boiling water after each
use. Dry the parts carefully using a soft cloth. Put
the parts together again when they are completely
dry and place them in a dry, sealed container or
disinfect them.
When cleaning, ensure that any residue is re-
moved. Never use any substances for cleaning
that could potentially be toxic if they came into
contact with the skin or mucous membranes, or if
they were swallowed or inhaled.
Use a soft, dry cloth and non-abrasive cleaning
products to clean the device.
Do not use any abrasive cleaning products and
never submerge the device in water.
Important
• Make sure that no water gets inside the device!
• Do not clean the device or atomiser set in the
dishwasher!
• Do not touch the device with wet hands when
it is plugged in and do not allow water to spray
on the device. Only operate the device if it is
completely dry.
• Do not spray any liquid into the vents! If liquid
penetrates the device this could cause dam-
age to the electrics or other nebuliser parts and
lead to a malfunction.
Condensation, hose care
Condensation may form in the hose depending on
the ambient conditions. It is essential to remove
the moisture to prevent bacterial growth and en-
sure proper treatment. To do so, proceed as fol-
lows:
• Disconnect the hose
• The hose must remain in the nebuliser
• Operate the nebuliser until the moisture is re-
moved by the air passing through.
• In the event of heavy contamination, replace
the hose.
, masks
13
, etc. must be
19
from the atomiser
8
.
6
Disinfection
Please carefully observe the points below when
disinfecting your atomiser and the atomiser set.
We recommend disinfecting the individual parts
on a daily basis after the last usage as a minimum
measure.
• First, clean the atomiser and atomiser set as
described in the "Cleaning" section.
14
• Place the disassembled atomiser
sal douche
masks
14 15
piece
in boiling water for 5 minutes. Con-
16
tact between the parts and the hot base of the
pan should be avoided.
• For disinfection immerse the compressed air
hose in a 2 % sodium hypochlorite (NaOCl)
solution for 15 minutes (i.e. a solution made
from the disinfecting agent Amuchina® or a
2 % sodium hypochlorite solution mixed by
your pharmacist). Afterwards rinse all parts
thoroughly with tap water, to fully remove any
possible residue of the disinfectant solution.
Important
Do not boil or autoclave the compressed air hose.
• Put the parts together again when they are
completely dry and place them in a dry, sealed
container.
Ensure that the parts are completely dried
after cleaning, otherwise the risk of bacterial
growth is increased.
Drying
• Place the individual parts on a clean, dry and
absorbent surface and leave them to dry com-
pletely (at least 4 hours).
Durability of materials
• As with any plastic parts, atomisers and the
rest of the atomiser set are a ected by a cer-
tain amount of wear and tear when used and
hygienically prepared on a frequent basis. Over
time, this can lead to a change in the aerosol,
which can have a negative e ect on the e -
ciency of the treatment. We therefore advise
.
9
you to replace the atomiser and the atomiser
set no later than one year if the parts are not
disinfected more than 360 times a year. Fre-
quent disinfection shortens the service life of
the parts.
• The following point should be observed when
selecting the cleaning product or disinfectant:
Only use a mild cleaning product or disinfect-
ant in the concentration and volume prescribed
by the manufacturer.
23
, mouthpiece
, the silicone
19
13
and the disassembled nose-
, na-
9