8. Press the On/O control
cation, place one finger over the cover hole
on the container.
During the application, breathe slowly and
evenly through your nose. Tilt your head
slightly in the opposite direction to the nostril
being treated so that the saline solution can
flow deep into the nose.
9. To pause or stop the application, take your finger
o the cover hole
23
10. Once your nose feels clear again, you can
stop the application. The application should
be stopped at the latest once the maximum
filling quantity has been used.
7.9 Cleaning
See "Cleaning and maintenance".
8. CHANGING THE FILTER
In normal operating conditions, the air filter must
be replaced after approx. 200 operating hours or
one year. Please check the air filter regularly (after
10-12 nebulisation procedures). Replace the used
filter if it is very dirty or clogged. If the filter has
become damp, it must also be exchanged for a
new filter.
Important
• Do not attempt to clean the used filter and re-
use it.
• Only use the manufacturer's original filter, oth-
erwise your nebuliser may become damaged
and su ciently e ective treatment cannot be
guaranteed.
• Do not repair or maintain the air filter while it
is in use.
• Never operate the device without a filter.
To replace the filter, proceed as follows:
Important
• First switch the device o and disconnect it
from the mains.
• Allow the device to cool down.
1. Remove the filter cap
If the filter remains in the device after the cap
has been removed, take the filter out of the
device, e.g. with tweezers or similar.
2. Re-insert the filter cap
. To start the appli-
3
on the saline container
upwards.
5
C
with a new filter.
5
3. Ensure that it is securely in place.
23
9. CLEANING AND MAINTE-
NANCE
Atomiser and atomiser set
Warning
Adhere to the following hygiene instructions to
.
22
avoid health risks.
• The atomiser
set are intended for multiple use. Please note
that di erent areas of application involve dif-
ferent requirements in terms of cleaning and
hygienic preparation.
• Do not clean the atomiser or the atomiser
set mechanically using a brush or similar
device, as this could cause irreparable
damage and it will mean that the best
treatment results can no longer be guar-
anteed.
• Please consult your doctor about the
additional requirements in terms of the
hygienic preparation needed (hand care,
handling of medication/inhalation solu-
tions) for high-risk groups (e.g. patients
with cystic fibrosis).
• Ensure thorough drying after each clean-
ing and disinfection process. Residual
moisture or wetness can represent an in-
creased risk of bacterial growth.
Preparation
• Immediately after each treatment, all parts of
the atomiser
used must be cleaned of medication residues
and contamination.
• To do this, dismantle the atomiser
individual parts.
• Remove the mouthpiece
nosepiece
• Disassemble the nosepiece if you have used it
with the comfort attachment.
• Dismantle the atomiser by twisting the top an-
ticlockwise against the medication container
.
12
• Remove the atomiser insert
cation container
• Remove the valve
atomiser
9
• Reassembly is carried out in reverse order.
22
and the rest of the atomiser
9
and the rest of the atomiser set
9
, mask
13
from the atomiser.
16
from the medi-
10
.
12
by pulling it out of the
11
.
into its
9
or
14 15