To disconnect the batteries on Front and Rear-wheel drive:
• Release the two handle screws under the mid of the seat
frame, that connect the seat frame with the seat module
interface (step 1) and release the lever (step 2) (Fig.6.6.1 -
6.6.2).
• Flip the seat frame backwards (Fig.6.3).
• Ensure the seat is secured with the seat stay in place, before
letting go, (Fig.6.7 Front - Fig.6.8 Rear).
• Once secured (Fig.6.4.2 - A), lift off the battery box lid, (Fig.
6.4.2 - B). You must make sure that the red seat stay is fully
latched onto the pin.
• Disconnect the 2 pin Anderson connectors from each battery
(Fig.6.5).
To reconnect the batteries reverse the above procedure.
WARNING:
Removing the batteries can lead to a change of the center point
of gravity. The result may be that the chair flips backwards
without the weight of the batteries. Sunrise advises to:
• Putting an obstacle behind the chair to secure the situation.
This can be a chest, or a chair, or etc.
• Try to lift the battery slightly first to check the effect of taking
out the weight.
24
Fig. 6.6.1
1
Fig. 6.6.2
Fig. 6.7
Fig. 6.8
FWD
2
RWD
2
1
FWD
RWD