This service information describes the connection of the IntelliBridge to the Fresenius Kabi Link+
Agilia infusion system.

Data communication

The IntelliBridge EC10 module supports RS232 serial data communication with the following
specifications for connection with the external device:
• Signaling speed
• Character bits
• Stop bits
RS232 parameters must be set and verified according to the Fresenius Kabi Link+ Agilia infusion
system service manual by authorized personnel.

Service Information

8
115200 Bd
8
1
• Parity generation
• Parity checking
8
None
None