NOTICE
Damage to the gas detector!
Harsh cleaning utensils (e.g. brushes), detergents and
solvents can destroy the dust and water filters.
► Only clean the gas detector with cold water and, where
necessary, a sponge.
For information on suitable cleaning agents and disin-
fectants and their specifications, see document
9100081 at www.draeger.com/IFU.
6
Device settings
Only trained and qualified personnel may change the device
settings.
For more details, refer to the technical manual.
6.1
Factory settings
Different settings can be selected to meet customer
requirements when ordering. The setting can be checked and
changed with the Dräger CC-Vision PC software.
Device settings via CC-Vision:
The changed parameter settings must be checked after
being transferred to the gas detector to ensure that the values
have been transferred correctly.
Parameters that cannot be viewed on the gas detector must
be read out and checked after being changed using the
Dräger CC-Vision PC software.
Device settings via a cloud solution:
In order to guarantee compliance with EN 50271, the
transmitted parameters must be verified when configuration
settings are changed.
Feature
Fresh air calibration without
1)
password
Bump test without password
2)
Hydrogen added signal
Life signal (acoustic)
Automatic distress signal
®
Bluetooth
Switch off is allowed
3)
Capture range
4)
LEL factor
5)
CH
(methane)
4
H
(hydrogen)
2
C
H
(propane)
3
8
Instructions for use
|
X-am
Setting
On
On
On
Off
Off
Off
On
On
4.4 Vol% (corresponds to
100 %LEL)
4.0 Vol% (corresponds to
100 %LEL)
1.7 Vol% (corresponds to
100 %LEL)
®
®
2600 / X-am
2800 / X-am
Feature
STEL
TWA
Configuration type alarm
thresholds
Alarm A1
Alarm A2
Expired bump test interval
Expired calibration interval
1) The fresh air calibration/zero calibration is not supported by the CO
channel of the infrared sensor and XXS O
2) With XXS H
HC activated and Ex channel of the DUAL IR Ex/CO
2
IR Ex activated.
3) The capture range can be activated or deactivated with the Dräger
CC-Vision PC software. The capture range is activated in measuring
mode ex works. The capture range is always deactivated in calibration
mode.
4) An LEL factor can be adapted to national regulations using the Dräger
CC-Vision PC software.
5) Notation in the gas detector: ch4
6.2
Device and sensor settings
Name:
Device settings:
Password(s)
Acoustic life signal
Switch-off mode
Short-term exposure limit
1)2)
(STEL)
Shift length (TWA)
Sensor settings:
A1 alarm:
Latching
Acknowledgeable
A2 alarm:
Acknowledgeable
®
5800
Device settings
Setting
STEL feature – inactive;
average duration =
15 minutes
TWA feature – inactive; aver-
age duration = 8 hours
ATEX compliant
Acknowledgeable; non-latch-
ing, pre-alarm, increasing
measured value (for O
sor, decreasing measured
value as well)
Not acknowledgeable; latch-
ing, main alarm, increasing
measured value (for O
sor, decreasing measured
value as well)
Channel warning
Channel warning
.
3
Area / setting
Numerical range (4-digit)
Yes / No
"Switch off is allowed" or
"Switch off not allowed" or
"Switching off not allowed
during A2"
0 - 60 (in minutes; setting for
exposure alarm)
60 - 1440 (in minutes; setting
3)
for exposure alarm)
On / Off
On / Off
On / Off
|
en
sen-
2
sen-
2
2
or
2
31