DS100A
Adult SpO
Sensor Reusable
2
Identification of a substance
that is contained or present
within the product or
packaging.
This device is not made with natural rubber latex or DEHP.
Directions for Use
Indications/Contraindications
The Nellcor™ Adult SpO
noninvasive arterial oxygen saturation and pulse rate monitoring are required for patients
weighing greater than 40 kg.
The DS100A is contraindicated for use on active patients or for prolonged use. It is not designed
for long-term monitoring. It must be moved every 4 hours (or more often, if indicated by
circulatory status and/or skin integrity) and reapplied to a different site. If long-term monitoring
is required, use an OxiMax™ oxygen sensor (MAXA, MAXAL, or MAXN) or an Oxisensor II sensor
(D25, D25L, or N25), depending on patient age and size, and instrument compatibility.
Instructions for Use
Use this sensor only with Nellcor instruments and instruments containing Nellcor oximetry,
or with instruments licensed to use Nellcor sensors (Nellcor-compatible instruments). This
sensor integrates Nellcor OxiMax technology into its design. When connected to an OxiMax-
enabled instrument, this sensor uses OxiMax technology to provide additional advanced
sensor performance features. Consult individual manufacturers for features and compatibility
of particular instruments and sensor models.
Reusable sensors may be used on the same site for a maximum of 4 hours, provided the site
is inspected routinely to ensure skin integrity and correct positioning. Because individual skin
condition affects the ability of the skin to tolerate sensor placement, it may be necessary to
change the sensor site more frequently with some patients.
To apply the DS100A:
1. Place an index finger over the sensor window of the DS100A with the finger tip against
the stop. (1)
Identification of a substance
that is not contained or
present within the product or
packaging.
Sensor, model DS100A, is indicated for use when continuous
2
IP22
3
Protected against access to hazard-
ous parts with a finger and protected
against vertically falling water drops
when sensor tilted up to 15°.