Optima Auto Link / English
Guidance and Manufacturer's Declaration- Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The user
of this device should make sure it is used in such an environment.
Immunity Test Levels
Professional
Basic EMC
healthcare
standard
facility
environment
Electrostatic
Discharge
±8 kV contact
(ESD)
± 2, 4, 8, 15 kV air
IEC61000-4-2
±2 kV, 100 kHz for power supply
Electrical fast
line
transient/ burst
±1 kV, 100 kHz for input/output
IEC61000-4-4
line
Surge
± 0.5, 1 kV (line to line)
IEC61000-4-5
± 0.5, 1, 2 kV (line to earth)
Voltage dips,
Voltage Dips:
short
i) 100% reduction for 0.5 period,
interruptions
ii) 100% reduction for 1 period,
and voltage
iii) 30% reduction for 25/30
variations on
period,
power supply
Voltage Interruptions:
input lines
100% reduction for 250/300
IEC61000-4-11
period
Power
frequency
(50/60Hz)
30 A/m
magnetic field
IEC61000-4-8
3 Vrms
0,15 MHz –
80 MHz
6 Vrms in ISM
Conducted RF
bands
IEC 61000-4-6
between 0,15
MHz and 80
MHz
80 % AM at 1
kHz
Compliance
HOME
Levels
HEALTHCARE
ENVIRONMENT
±8 kV contact
± 2, 4, 8, 15
kV air
±2 kV, 100
kHz for
power supply
line
± 0.5, 1 kV
(line to line)
30 A/m
30 A/m
3 Vrms
0,15 MHz –
3 Vrms
80 MHz
0,15 MHz – 80
6 Vrms in
MHz
ISM and
6 Vrms in ISM
amateur
and amateur radio
radio bands
bands between
between
0,15 MHz and 80
0,15 MHz
MHz
and 80 MHz
80 % AM at 1 kHz
80 % AM at 1
kHz
Electromagnetic
Environment-Guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30 %.
Mains power quality should
be that of atypical
commercial or hospital
environment
Mains power quality should
be that of atypical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
this device requires
230 Vac
continued operation during
power mains interruptions,
it is recommended that the
device be powered from an
uninterruptible power
supply or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Portable and mobile RF
communications equipment
should be used no closer to
any part of this device,
including cables, than there
commended separation
distance calculated from
the equation applicable to
the frequency of the
transmitter.
Recommended separation
25