Optima Auto Link / English
C. Tuning
Original setting
NOTE: The pressure in the air cells can be manually adjusted for patient
comfort by entering the tuning mode in the lower right corner of the LCD
screen.
NOTE: Slide one hand underneath the air cell where the patient's buttock
locates to ensure that the given air pressure is appropriate. Always leave a
space for about 2.54 cm (1 inch) between the patient and the cell to prevent
bottoming-out.
4.2. Internet of Things Service (IoT)
The Optima Auto Link is an IoT-technology-enabled device. This IoT device streams the pump
or mattresses' real-time conditions via a cloud-based platform named KeepWell.
For the IoT device, the LCD displays a signal icon that shows the
current strength.
For more detail, please refer to the manual of the cloud-based
platform: "KeepWell—streamline operation, logistics, and customer service for your business."
NOTE: Use this device in places where 4G LTE NETWORK coverage is good
to ensure optimal IoT services.
NOTE: Consult with the service provider or the manufacturer for the
cellular communication service provider.
4.3. Emergency CPR Operation
CPR knob is located at the patient's left-hand side of the
mattress near the head area. When there is a need to
perform CPR on the patient, quickly turn the CPR knob
to release air from mattress. The quick connector on the
pump unit can be disconnected for even faster deflation.
Decrease mattress
pressure
Increase mattress
pressure