Skin dose levels according to IEC 60601-2-54
Dosimetric information:
Skin dose values have been measured according to IEC 60601-2-54 chapter
203.5.2.4.101. Selected settings: SID: 115cm; Filter 1 mm AL and 0.1mm Cu;
Pulse repetition frequency (duration) per DTS 5.2 seconds; Focal spot and
angle depends on the body part. Measuring setup includes a dosimeter, which
is positioned on top of an empty polymethyl-methacrylate (PMMA) phantom
according to IEC 60601-2-54 chapter 203.5.2.4.5.102.
The user documentation of this product contains a set of skin dose value
measurements. Refer to document "Exposure Techniques for pediatric and
adult use with DR 600".