Diana II
Battery charge display and charging
Caution!
• The product cannot be used while charging.
• During charging, the lifting function is
locked.
• Do not use the lift for 1 hour after charging.
This will prolong the life of the battery.
• The battery needs to be charged continuous-
ly for at least 24 hour,
in the following situations:
a) Before using the lift for the first time.
b) Before being placed in storage (up to 3
months) without a power supply connec-
tion.
c) Immediately after a storage period.
Battery charge indicator
The battery charge indicator has three status
indications. The battery indicators are either
yellow or green until switched off (2 minutes
after use).
Battery status 1:
The battery is full – no recharging required
(100-50%): the third indicator is green.
Battery status 2:
The battery is okay (50-25%) – recharging
required: the second indicator is yellow.
Battery status 3:
The battery is low (less than 25%) – recharging
required: the first indicator is yellow and a buzzer
sounds whenever a button is pressed.
Battery status 4:
The battery requires recharging. Some functions
of the lift are lost and it is only possible to lower
the boom. When in continued use, the indicator
flashes and an audio signal is emitted.
18
www.humancaregroup.com
The battery charge is indicated on the
hand control.
Battery level indicators on the hand
control