Data Logging
7
Data Logging
Fig. 15: Data Logging
50
NOTE!
Data logging should be done with the DC adaptor
connected, to conserve the batteries. When the
DC adaptor is used, it will cut off the power supply
from the batteries.
The meter can automatically record readings of
CO
, CO, temperature and humidity for long time
2
environment monitoring. The memory capacity is
6.000 records with all values measured at a cer-
tain time, thus 24.000 readings.
The user can set up a sampling rate from 1 se-
cond to 4:59:59 hours (see chapter 11.4).
After the sampling rate is selected, press
START/ESC for two seconds in the normal
measuring mode.
The REC icon flashes to indicate the logging sta-
tus. The LCS main display shows the real time
value. In the lower part of the screen, date and
time are displayed in turns.
To terminate the data logging, press the
START/ESC key for two seconds, until the
REC icon stops flashing.
NOTE!
When the user presses START/ESC to start the
logging again, the previous logging records will be
overwritten. Export the Logged records to the PC
to save them (see chapter 14).
Press MODE/RECALL to switch between the
different readings of the record. This is even
possible during the logging process.
NOTE!
During the logging process, MIN/MAX/AVG, man-
ual record, hold and recall functions are disabled.
During the logging process, the meter cannot be
switched off.