Storage Procedure
To minimize clogging and ensure quick response time, the glass bulb and the junction
should be kept hydrated.
Add a few drops of MA9015 Storage solution (not included with the bench meter) to the
protective cap. Replace the storage cap when the electrode is not in use.
Note: Never store the electrode in distilled or deionized water.
Regular Maintenance
1. Inspect the electrode for any scratches or cracks. If any present, replace the electrode.
2. Inspect the cable. The connection cable and insulation must be intact.
3. Connectors should be clean and dry.
4. Rinse off salt deposits with water.
5. Follow storage procedure above.
For refillable electrodes:
Refill the electrode with fresh electrolyte solution (see the electrode's specifications
to select the correct refilling solution). Keep the electrode upright for 1 hour. Follow the
storage procedure above.
Cleaning Procedure
General: Soak the electrode in MA9016 Electrode cleaning solution for approximately 30
minutes (not included with bench meters).
IMPORTANT: After performing any of the cleaning procedures, rinse the electrode
thoroughly with distilled water and soak in MA9015 Storage solution for at least 1 hour
before taking measurements.
ELECTRODE CONDITION & MAINTENANCE