pH VERSUS TEMPERATURE CHART
TEMP
pH VALUES
°C
°F
MA9004
MA9007
0
32
4.01
5
41
4.00
10
50
4.00
15
59
4.00
20
68
4.00
25
77
4.01
30
86
4.02
35
95
4.03
40
104
4.04
45
113
4.05
50
122
4.06
55
131
4.07
60
140
4.09
65
149
4.11
70
158
4.12
OPTIONAL ACCESSORIES
M10007B pH7.01 buffer solution, 25x20 mL sachet
M10004B pH4.01 buffer solution, 25x20 mL sachet
M10010B pH10.01 buffer solution, 25x20 mL sachet
M10031B 1413 S/cm calibration solution, 25x20 mL sachet
M10032B 1382 ppm calibration solution, 25x20 mL sachet
M10442B 1500 ppm calibration solution, 25x20 mL sachet
MA9015
Electrode storage solution, 220 mL bottle
SE-600
pH/EC/TDS spare probe with 1m cable
SPECIFICATIONS:
RANGE
0.00 to 14.00 pH (MW802)
0 to 1990 ppm; 0 to 1990 µS/cm (MW801)
0 to 4000 ppm; 0.00 to 6.00 mS/cm (MW802)
RESOLUTION
10 ppm (MW801/MW802)
ACCURACY (@25°C)
2% FS (EC/TDS) (MW801, MW802)
TDS FACTOR
TEMPERATURE COMPENSATION
CALIBRATION
MA9010
PROBE
7.13
10.32
ENVIRONMENT
7.10
10.24
BATTERY TYPE
7.07
10.18
BATTERY LIFE
7.04
10.12
DIMENSIONS
7.03
10.06
WEIGHT
7.01
10.01
CERTIFICATION
7.00
9.96
Milwaukee Instruments conform to the CE
6.99
9.92
European Directives.
6.98
9.85
Disposal of Electrical & Electronic Equipment.
6.98
9.85
Do not treat this product as household waste. Hand
6.98
9.82
it over to the appropriate collection point for the
recycling of electrical and electronic equipment.
6.98
9.79
Disposal of waste batteries. This product
6.98
9.77
contains batteries. Do not dispose of them with
6.99
9.76
other household waste. Hand them over to the appropriate
6.99
9.75
collection point for recycling.
Please note: proper product and battery disposal
prevents potential negative consequences for human
health and the environment. For detailed information,
contact your local household waste disposal service or
go to www.milwaukeeinstruments.com (USA & CAN) or
www.milwaukeeinst.com.
RECOMMENDATION
Before using this product, make sure it is entirely suitable for
your specific application and for the environment in which it is
used. Any modification introduced by the user to the supplied
equipment may compromise the meter's performance. For
your and the meter's safety do not use or store the meter
in hazardous environment. To avoid damage or burn, do not
perform any measurement in microwave ovens.
0.0 to 14.0 pH (MW801)
WARRANTY
These instruments are warranted against defects in materials
and manufacturing for a period of 2 years from the date of
0.1 pH (MW801)
purchase. This warranty is limited to repair or free of charge
0.10 pH (MW802)
replacement if the instrument cannot be repaired. Damage
due to accidents, misuse, tampering or lack of prescribed
10 µS/cm (MW801)
maintenance is not covered by warranty. If service is required,
0.01 mS/cm (MW802)
contact your local Milwaukee Instruments Technical Service.
±0.2 pH (MW801)
If the repair is not covered by the warranty, you will be notified
±0.20 (MW802)
of the charges incurred. When shipping any meter, make sure
it is properly packaged for complete protection.
0.5 (MW801)
0.68 (MW802)
ISTMW802 07/20
Automatic, 0 to 50°C
Manual, 1-point
SE-600 combination probe
0 to 50°C; 95% RH max.
1 x 9V
approx. 150 hours of use
143 x 80 x 32 mm
220 g (with battery)
SATISFACTION
GUARANTEED
USER MANUAL
MW801, MW802
PRO pH/EC/TDS
Portable Meters
PPM
pH
mS/cm
PRO pH/EC/TDS Meter
milwaukeeinstruments.com (USA & CAN)
milwaukeeinst.com