6. SETUP
Setup mode allows parameters that are not directly related to the measurement to be
viewed and modified.
1. Press SETUP to enter in Setup mode.
2. Use  keys to select required parameter.
3. Press the CAL/EDIT to enter Edit mode.
4. If various options are available, use RANGE/ to select desired option.
5. Use  keys to select or change parameter values.
6. Press GLP/ACCEPT to confirm or press CAL/EDIT to exit without changing.
7. Once in Edit mode, below detailed parameters can be set or viewed.
If needed, use ESC to exit Setup mode.
6.1. GENERAL SETUP
pH Information
Use  keys to select whether the pH buffer calibration information is enabled or not:
on – enable (default), off – disable.
When enabled, the electrode segment displays the electrode condition (MW151 only).
Calibration Expired Warning
Use  keys to select the number of days since last calibration has elapsed: 1 to 7 days
(default), or off.