2. Place, align and tighten the pump adapter on the thread
port on the top cover. Check that the pump adapter is
mounted correctly. Avoid bending the pump adapter. The
gas detector automatically switches to pump mode as
soon as the pump adapter is mounted.
The leak test starts automatically.
3. When the leak test is displayed, the suction inlet on the
probe or hose closes within 60 s and remains closed until
the leak test is complete.
4. Release the inlet opening.
● Leak test successful: Measurement starts.
● Leak test failed: Inspect the accessories and the pump
adapter and then repeat the leak test.
5. Observe the flushing times. For further information, see
the following chapter: "Special features when measuring
with the pump", page 34
6. Remove the pump adapter.
7. After completing the measurement, make sure that the
seal on the pump adapter is clean and that there are no
chips of metal on the seal.
Plug the protective cap onto the pump adapter to protect
the seal from damage and being deformed.
4.7.2 Alarms
In the event of an alarm, corresponding displays, the optical
alarm, vibration alarm as well as, if necessary, the acoustic
alarm are activated (configurable). For further information,
see the following chapter: "Alarm settings (factory
setting)", page 42
To acknowledge an alarm:
1. Select
.
4.7.3 Special state
The life signal is disabled during a special state. Special
states are displayed by the following visual signals:
– Yellow LED flashing – 'warm-up 1' special state
– Yellow LED continuously illuminated – general special
state
No alarms are issued during a special state.
Exception: The calibration adapter is mounted in measuring
mode. In this case, alarms continued to be issued as long as
the measurement gas can reach the sensors.
The special state is exited by eliminating the potential error, by
switching to normal measuring mode in a correctly functioning
gas detector or it will switch off itself after approx. 1 minute.
WARNING
Incorrect measurement!
A mounted calibration adapter blocks free gas diffusion to the
sensors. Correct measured values and alarms can no longer
be guaranteed.
► It is imperative to actively bump the sensors (e.g. test gas
cylinder with pressure reducer, flow 0.5 L/min).
4.7.4 Protection alarm
The protection alarm protects the CatEx sensor.
Instructions for use
|
X-am
®
3500
If the measuring range is exceeded significantly at the CatEx
channel (very high concentration of flammable substances), a
protection alarm is triggered. This CatEx protection alarm can
be acknowledged by switching the gas detector off and then
on again in fresh air.
If the gas detector cannot be switched off because the A2
alarm is active and the switch-off mode in the CC-Vision is set
to "Switching off not allowed during A2", remove the power
pack or place the gas detector in the charging cradle and
allow it to switch off automatically.
For more details, refer to the technical manual.
4.7.5 Deleting (application) peaks
1. Select
in measuring mode.
2. Select Clear app. peak and confirm the dialog.
The function must be activated in the quick menu.
Alternatively, this function can also be called via the menu.
4.8
Calling the Quick Menu
The Dräger CC-Vision software can be used to save up to 6
preferred features.
The following features are default settings:
– Device information
– Night mode
– Display shift peaks
– App.peak value
– Delete app. peaks
– Messages
To open the quick menu:
1. Select
in measuring mode.
2. Select and confirm the desired feature.
4.9
Opening information
1. Select
> Info in measuring mode.
The following options are available:
Option
Messages
Device information
|
Operation
Description
The pending warnings and
errors are displayed. For a
description of the messages
and remedial measures, see
the technical manual.
Device information and infor-
mation about the Bluetooth®
module (optional, X-am 8000
only) is displayed (e.g. MAC
address, serial number, firm-
ware version etc.).
en
33