E1
EVACUATION PIPE
The HP unit contains evacuation pipes, which are exhaust pipes from the intermediate pressure
safety valves.
E2
EVACUATION PIPE
The Stabilizer contains evacuation pipes which are exhaust pipes from distribution pressure
safety valves.
O1
OUTLET OF HP UNIT
The HP unit have one outlet with thread connection. On the outlet of HP unit it is possible to
mount shut off valve and outlet pipe. The Shut off valve and outlet pipe are included only in
DUPLEX (manifold with HP unit, Stabilizer and Alarm).
O2 OUTLET OF STABILIZER
The Stabilizer has one outlet connection. The connection is with swivel nut and internal thread
connection.
P9 CONNECTION
The Stabilizer contains one connection which can be used as outlet of the Stabilizer. The con-
nection is with swivel nut and internal thread connections.
P1, P2
CONNECTION
The HP unit contains two connections which are used during service and repair, for example
testing safety valves.
P3
CONNECTION
HP unit contains one connection which is used during service and repair, for example digital
pressure device.
P7, P8
CONNECTION
The Stabilizer contains two connections which are used during service and repair, for example
testing safety valves.
V11
V13
H1
The MC80 semiautomatic medical manifold can be delivered with HP accessories. The ac-
cessories contain parts for connecting gas cylinders. Basic parts of accessories are in picture
bellow.
P11, P12
COLLECTING PIPES
Collecting pipes are used for connecting gas cylinders to the manifold. Two equal parts one left
and one right of these pipes should be included to the HP unit.
P11
P12
9/170
EN
V12
V14
H2