5. PRODUCT DESCRIPTION

The manifolds covered by these instructions are designed to allow equal numbers of cylinders
to be manifold together to give an operating bank and a reserve bank. The operating bank will
deliver gas to the manifold pressure regulator until the cylinders are exhausted. At that point the
supply will switch to the reserve bank and the exhausted bank can be replenished. The object
is to give uninterrupted gas supply.
The MC80 semiautomatic (DUPLEX) medical manifold contains three parts: HP UNIT, STABI-
LIZER. According to product variant you obtain one or all parts of MC80.
EN
HP UNIT
STABILIZER
HP UNIT
The manifold MC80 contains High pressure unit. More info about High pressure unit is men-
tioned below.
STABILIZER
The manifold MC80 contains Stabilizer. More info about Stabilizer is mentioned below.
GAS ALARM
The manifold MC80 contains Gas Alarm. For more info see Instruction for use for Gas Alarm.
I1, I2
INLET CONNECTION
The HP unit contains left and right side inlet collection pipes. There are thread connections that
fi ts to the high pressure hoses.
I3
INLET OF STABILIZER
The Stabilizer contains one inlet connections. There is swivel nut with internal thread connec-
tion.
P5, P6
CONNECTION
The Stabilizer contains two connections which can be used as inlet of the Stabilizer, for example
for LOX connection.The connections are with swivel nuts and internal thread connections.
F1, F2
FILTERS
The manifold contains two 100 microns high pressure fi lters with integrated non-return valves
which are installed on the inlet of the HP unit. These fi lters are easy to maintain.
V1, V2
INLET SHUT OFF VALVES
The manifold contains two inlet shut off valves which are for opening and closing main high
pressure supply from the gas cylinders to the left and right side of the HP unit.
V3, V4
SHUT OFF VALVES
The HP unit contains two shut off valves. The shut off valves are used together with V1 and V2
during service and repair.
V5 SHUT OFF VALVE
The HP unit contains one shut off valve which is used during service and repair of the pilot regu-
lator. The Shut off valve is used for closing pilot pressure for high pressure regulators.
6/170