11 Additional functions
11.1.3 Factory default setting
The continuously displayed CAL symbol indicates that an adjustment has already taken place. How-
ever, the factory default setting should be restored.
1. Entering CAL mode
a.
b.
2. Restoring the factory default setting
a.
70
Operating manual
a.
With the instrument powered on, press and hold the
> 3 s
CLEAR button for more than 3 seconds until the follow-
ing modes are repeatedly shown one after another in
the display:
CAL — GLP — APO — dP
b.
When CAL appears in the display, release the CLEAR
button. CAL blinks and the adjustment value of the pre-
vious adjustment appears.
Press the On/Off button and the Pause button at the same
time to restore the factory default setting. The CAL symbol
disappears.
!