This service information describes the connection of the IntelliBridge to the Dräger Carina ventilator.
Data Communication
The IntelliBridge EC10 module supports RS232 serial data communication with the pre-configured
communication settings of the Dräger Carina. No further configuration of the connection parameters
is needed.
Instruction for Connection
Connect the IntelliBridge EC5 D-Sub 9-pin connector to the RS232 connector of the Dräger Carina
ventilator as shown in the photo below. Make sure to tighten the screws of the connector properly.
IntelliBridge EC5 connection to RS232 ports
1
2
3
References
For further service information, refer to the IntelliVue Patient Monitor Service Guides.
Service Information
7
1
Dräger Carina ventilator - RS232 (COM) port
IntelliBridge EC5 ID module
EC5 device label: "Dräger Carina"
2
3
7