This service information describes the connection of the IntelliBridge to the Dräger Zeus
Empowered Anesthesia Workstation.
Data communication
The IntelliBridge EC10 module supports RS232 serial data communication with the following
specifications for connection with the external device:
• Baud rate:
• Word length:
RS232 parameters must be set and verified according to the Dräger Zeus IE Anesthesia Workstation
service manual by authorized personnel.
Instruction for connection
Connect the IntelliBridge EC5 D-Sub 9-pin connector to the RS232 connector of the Dräger Zeus IE
Anesthesia Workstation as shown below. Make sure to tighten the screws of the connector properly.
IntelliBridge EC5 connection to RS232 ports
1
2
3
References
For further service information, refer to the patient monitor service guide.
Service Information
7
38400
8 bits
1
Dräger Zeus IE Anesthesia Workstation - RS232 (Medibus 1) port
IntelliBridge EC5 ID module
EC5 device label: "Dräger Zeus IE"
• Stop bits:
• Parity/Checking:
3
7
®
®
Infinity
1
Even
2
31