MicroRAE Gebruikershandleiding
ONLY THE COMBUSTIBLE GAS DETECTION PORTION OF THIS INSTRUMENT HAS
BEEN ASSESSED FOR PERFORMANCE.
ALLEEN DE PRESTATIES VAN HET DEEL VAN DIT INSTRUMENT DAT BRANDBAAR
GAS DETECTEERT, ZIJN BEOORDEELD.
CAUTION: BEFORE EACH DAY'S USAGE, SENSITIVITY OF THE COMBUSTIBLE GAS
SENSOR MUST BE TESTED ON A KNOWN CONCENTRATION OF METHANE GAS
EQUIVALENT TO 20 TO 50% OF FULL-SCALE CONCENTRATION. ACCURACY MUST BE
WITHIN 0 AND +20% OF ACTUAL. ACCURACY MAY BE CORRECTED BY CALIBRATION
PROCEDURE.
VOORZICHTIG: ELKE DAG WAAROP HET INSTRUMENT WORDT GEBRUIKT, MOET
EERST DE GEVOELIGHEID VAN DE SENSOR VOOR BRANDBARE GASSEN WORDEN
GETEST OP EEN BEKENDE CONCENTRATIE METHAANGAS DIE GELIJK IS AAN 20 TOT
50% VAN DE VOLLE SCHAALCONCENTRATIE. DE NAUWKEURIGHEID MOET TUSSEN 0
EN +20% VAN DE EFFECTIEVE WAARDE LIGGEN. DE NAUWKEURIGHEID KAN
WORDEN VERBETERD MET BEHULP VAN EEN KALIBRATIEPROCEDURE.
CAUTION: HIGH OFF-SCALE READINGS MAY INDICATE AN EXPLOSIVE
CONCENTRATION.
VOORZICHTIG: HOGE METINGEN BUITEN DE SCHAAL KUNNEN WIJZEN OP EEN
EXPLOSIEVE CONCENTRATIE.
WAARSCHUWINGEN
5