50 ...1000 Hz
+ (1.5 % + 7 d)
+ (1.5 % + 7 d)
accuracy
fied from 5 % of
All manuals and user guides at all-guides.com
Frequency (elec-
40.00 Hz ...10 KHz 0.1 Hz ...0.001 kHz
trical)
Sensitivity: 1 V rms
Duty cycle
0.1 ... 99.90 %
Pulse width: 100 μs...100 ms
Frequency: 5 Hz...150 kHz
Temperature
-50
C...1000
o
(type K)
-58
F....1832
o
4-20 mA %
-25 %...125 %
0 mA = -25 %, 4 mA = 0 %, 2- mA = 100 %, 24 mA = 125 %
Mega Ohms
Terminal
Range
voltage
125 V
0.125~4.000 MΩ
(0 % ~ +10 %)
4.001~40.00 MΩ
40.01~400.0 MΩ
400.1~4000 MΩ
250 V
0.250~4.000 MΩ
(0 % ~ +10 %)
4.001~40.00 MΩ
40.01~400.0 MΩ
400.1~4000 MΩ
500 V
0.500~4.000 MΩ
(0 % ~ +10 %)
4.001~40.00 MΩ
40.01~400.0 MΩ
400.1~4000 MΩ
1000 V
1.000~4.000 MΩ
(0 % ~ +10 %)
4.001~40.00 MΩ
40.01~400.0 MΩ
400.1~4000 MΩ
Note: Accuracy specifications consist of two elements:
• (% reading) – This is the accuracy of the measurement circuit
• (+ digits) – This is the accuracy of the analog to digital converter
+ (0.5 %)
0.01 %
+ (1.2 % + 2 d)
C
0.1
C
+ (1.0 % + 2.5
o
o
F
0.1
F
o
o
+ (1.0 % + 4.5
(excl. probe
accuracy)
0.01 %
+ (50 d)
Resolu-
Accuracy
Test
tion
current
0.001 MΩ
+ (2.0 % + 10)
0.01 MΩ
+ (2.0 % + 10)
1 mA @
100 kΩ
0.1 MΩ
+ (4.0 % + 5)
1 MΩ
+ (5.0 % + 5)
0.001 MΩ
+ (2.0 % + 10)
0.01 MΩ
+ (2.0 % + 10)
1 mA @
250 kΩ
0.1 MΩ
+ (3.0 % + 5)
1 MΩ
+ (4.0 % + 5)
0.001 MΩ
+ (2.0 % + 10)
0.01 MΩ
+ (2.0 % + 10)
1 mA @
500 kΩ
0.1 MΩ
+ (2.0 % + 5)
1 MΩ
+ (4.0 % + 5)
0.001 MΩ
+ (3.0 % + 10)
1 mA
0.01 MΩ
+ (2.0 % + 10)
@ 1000
0.1 MΩ
+ (2.0 % + 5)
kΩ
1 MΩ
+ (4.0 % + 5)
www.nieaf-smitt.com
C)
o
F)
o
Short
circuit
current
< 1 mA
< 1 mA
< 1 mA
< 1 mA
Nieaf-Smitt is a brand name of Mors Smitt
63