Appendix
Tolerances
Refrigerators (RR)
BioMidi
425 (Solid door)
425 (Glass door)
625 (Solid door)
625 (Glass door)
Refrigerators with extended refrigeration (ER)
BioPlus
500 (Solid door)
500 (Glass door)
600D/600W (Solid door)
600D/600W (Glass door)
660D/660W (Solid door)
660D/660W (Glass door)
930 (Solid door)
1270/1400 (Solid door)
1270/1400 (Glass door)
*):
The temperature span between the initial temperature and the end temperature in the hold-over test P-13,14.
**):
Please contact your local distributor for current information.
***): 90° opening 1 minute
Ambient Temperature
RR/ER
Name:
Conducted by:
Inspected/verified by:
156
Door
Permissable
opening
deviation
recovery
time
(minutes)
***
3 min.
–
4 min.
–
+/- 3K
3 min.
–
4 min.
–
3 min.
–
4 min.
–
3 min.
–
4 min.
–
+/- 2K
3 min.
–
4 min.
–
5 min.
–
5 min.
–
7 min.
–
Setpoint temperature
+25 °C
Model: ________________________
Pull-down
Permissable
(minutes)
deviation
20 min.
–
25 min.
–
20 min.
–
25 min.
–
22 min.
–
28 min.
–
20 min.
–
25 min.
–
20 min.
–
25 min.
–
22 min.
–
23 min.
–
29 min.
–
+5 °C
Signature:
Hold-over
Hold-over
Permissable
range*
63 min.
37 min.
-5 °C → 10 °C
63 min.
37 min.
72 min.
42 min.
70 min.
41 min.
-5 °C → 10 °C
70 min.
41 min.
65 min.
78 min.
45 min.
Approved
Date:
(Yes/No):
SN: ________________________
deviation
–
–
–
–
–
–
–
–
–
–
–
–
–