Installation Qualification – IQ
ID
Description of installation
Ensure the cabinet is installed in-
I-1
doors.
Ensure the cabinet is installed in a dry
I-2
and sufficiently ventilated area.
Ensure the cabinet is not in direct
I-3
contact with sunlight or other heat
sources.
Ensure that the ambient operating
I-4
temperature is within the allowed
range.
Ensure that the cabinet is not
I-5
installed in a chloric/acidic environ-
ment.
Ensure that the protective film on the
I-6
cabinet is removed.
Ensure that the cabinet is cleaned
I-7
with a mild soap solution
Ensure that the cabinet has stood
I-8
upright for 24 hours if the cabinet has
been laying down.
Ensure that the cabinet is levelled if it
I-9
is equipped with legs.
Ensure a level surface if the cabinet is
I-10
equipped with wheels/castors.
If equipped with wheels/castors: En-
I-11
sure wheels/castors are locked after
positioning
If equipped with drawers and/or
I-12
glass door: Ensure that tilt-bracket is
mounted.
Ensure a distance of 15-75 mm
I-13
between cabinet and back wall.
Ensure that there is minimum a gap
I-14
of 30 mm between cabinets.
Ensure that the upper part of the
I-15
cabinet is not covered.
Ensure that electrical appliances are
I-16
not being used inside the cabinet.
Ensure connection from voltage-free
I-17
contact to external monitoring system
140
Comply
Reference
in manual
Yes
No
page 10
page 10
page 10
page 10
page 10
page 10
page 10
page 10
page 11
page 11
page 11
page 12
page 13
page 14
page 14
page 14
page 15
Model: ________________________
Attachment
Notes
SN: ________________________