Operation Qualification – OQ
ID
Description of operation
Turn on the cabinet – Display test
O-1
(software version and variant).
O-2
Set/adjust set-point temperature.
Set/adjust LhL – Upper alarm limit
O-3
(local).
Set/adjust LLL – Lower alarm limit
O-4
(local).
Set/adjust Lhd – Delay of the upper
O-5
alarm limit (local).
Set/adjust LLd – Delay of the lower
O-6
alarm limit (local).
Activate / deactivate dA – Door alarm
O-7
(local).
Set/adjust dAd – Delay of the door
O-8
alarm (local).
Activate / deactivate BU – Acoustic
O-9
alarms (local).
Set/adjust EhL – Upper alarm limit
O-10
(external).
Set/adjust ELL – Lower alarm limit
O-11
(external).
Set/adjust Ehd – Delay of the upper
O-12
alarm limit (external).
Set/adjust ELd – Delay of the lower
O-13
alarm limit (external).
Activate / deactivate dA – Door alarm
O-14
(external)
Set/adjust dAd – Delay of the door
O-15
alarm (external).
Activate / deactivate BU – Acoustic
O-16
external alarms.
Set/adjust defrost cycles (dEF) per 24
O-17
hours (factory setting: 4).
Select reference sensor for the
O-18
display (dPS) (A or E).
142
Comply
Reference
in manual
Yes
No
page 20
page 20
page 24
page 24
page 25
page 25
page 26
page 26
page 27
page 28
page 28
page 29
page 29
page 30
page 30
page 31
page 35
page 36
Model: ________________________
Attachment
Notes
SN: ________________________