Appendix
Tolerances
FREEZERS (RF)
BioPlus
500
600D/600W
660D/660W
930
1270/1400
EXTENDED FREEZERS (EF)
BioMidi
425
BioPlus
600W/660W
*):
The temperature span between the initial temperature and the end temperature in the hold-over test P-13,14.
**):
Please contact your local distributor for current information.
***): 90° opening 1 minute
Ambient Temperature
RF
EF (425)
EF (600W/660W)
Name:
Conducted by:
Inspected/verified by:
Door
Permissable
opening
deviation
recovery
time
(minutes)
***
7 min.
–
7 min.
–
+/- 5K
7 min.
–
–
–
10 min.
–
+/- 9K
40 min.
–
+/- 10K
30 min.
–
Setpoint temperature
+25 °C
+25 °C
+25 °C
Model: ________________________
Pull-down
Permissable
(minutes)
deviation
45 min.
–
42 min.
–
42 min.
–
-20 °C → 10 °C
76 min.
–
45 min.
–
107 min.
–
-40°C → 10 °C
57 min.
–
-35°C → 10 °C
-20 °C
-40 °C
-35 °C
Signature:
b i o l i n e
Hold-over
Hold-over
Permissable
range*
deviation
55 min.
55 min.
55 min.
63 min.
58 min.
108 min.
170 min.
Approved
Date:
(Yes/No):
SN: ________________________
–
–
–
–
–
–
–
157