With the X-am 8000 module: Loosen the screws at the
hose coupling and remove the hose coupling with the
hoses. Check that the seals of the hose coupling are intact,
replace the hose coupling (2) if required.
4. Replace the sealing insert.
5. With the Pac/X-am 125 module: Fit the hoses to the new
sealing insert (note the arrows on the sealing insert and the
hose).
With the X-am 8000 module: Fit the hose coupling with the
hoses on the new sealing insert and tighten the two
screws.
6. Squeeze the external locking lugs in and insert the sealing
insert into the module cover. The locking lugs must click
into place.
7. Check that the sealing insert is fitted correctly in the module
cover.
6.7
Changing the fresh air filter
NOTICE
i
i
With regular use and depending on the conditions of
use, the fresh air filter should be changed typically
every 2 months.
1. Unscrew the old fresh air filter.
2. Screw in the new fresh air filter.
6.8
Calibrating the touchscreen
1. At system start hold the function key until the calibration
screen is displayed.
2. Press each of the 5 position markers displayed in succession.
6.9
Cleaning
CAUTION
!
Impairment and wear of the system!
Heavy contamination or dust accumulation may impair
the function of the system and lead to increased wear
of individual system components.
Clean the system at regular intervals.
Abrasive
cleaning
cleaning agents and cleaning solvents can destroy the
fresh air filter.
The device does not need any special care.
If very dirty, the equipment can be carefully wiped down
with a damp cloth.
Carefully dry the device using a cloth.
7
Disposal
This product must not be disposed of as municipal
waste. This is indicated by the adjacent icon.
You can return this product to Dräger free of charge.
For
information
marketing organisations and Dräger.
Dräger X-dock 5300/6300/6600
implements
(brushes,
please
contact
the
national
8
Technical data
Dimensions (H x W x D):
Master
Pac / X-am 125
module
X-am 8000 module
Weight:
Master
Pac / X-am 125
module
X-am 8000 module
X-am 8000+ module approx. 1445 g
Ambient conditions:
During operation
During storage
During charging
(X-am 8000+
module only)
Gas connections:
X-dock 5300/6300
X-dock 6600
Inlet pressure:
for the measured gas 0.5 bar ±20 %
for compressed air
Power supply:
etc.),
Connections:
Serial no. (year of
manufacture):
approx. 3.54 x 5.70 x 9.84 inches
(120 x 130 x 250 mm)
approx. 3.54 x 5.70 x 9.84 inches
(90 x 145 x 250 mm)
122 x 145 x 250 mm
approx. 4.02 pounds
(48.23 ounces; 1500 g)
approx. 2.57 pounds
(30.86 ounces; 960 g)
approx. 1225 g
32 °F to +104 °F (0 °C to +40 °C)
-4 °F to +122 °F (-20 °C to +50 °C)
700 to 1300 hPa
max. 95 % relative humidity
0 °C to +35 °C (outside of this
temperature range charging is possible
only to a limited extent)
1x fresh air connection
1x compressed air inlet
1x exhaust outlet
3x gas inlets
6x gas inlets
0.5 bar ±20 %
11 V - 28 V DC, 6.25 A
3x USB 2.0 standard A connection,
(host, cable <3 m)
1x USB 2.0 mini-B connection,
(device, cable <3 m)
1x Ethernet port RJ45
Data rate10/100 Mbit
Composition of the serial numbers: The
third letter of the serial number
indicates the year of manufacture:
M = 2019, N = 2020, P = 2021,
R = 2022, S = 2023, T = 2024,
U = 2025, W = 2026, X = 2027,
Y = 2028, Z = 2029 etc. (letters G, I, O,
Q are omitted)
Example: Serial number ARMB-0001:
The third letter is M, which means that
the device was manufactured in 2019.