1. Enter volume of gas cylinder.
2. Set type of volume:
Compr.= cylinder volume
Relaxed = gas volume (cylinder volume x pressure)
3. If required, refresh test gas pressure.
4. Select Hose > 2m if the length of the test gas hose
exceeds 2 m.
NOTICE
i
i
Dräger does not recommend exceeding a hose length
of 10 m for the gas supply hoses.
5. If required, configure other gas inlets in the same way.
State of gas pressure:
Gas pressure Meaning
Display
<0.4 bar
0.4 - 0.6 bar
>0.6 bar
Gas cylinder level indicator:
Display
Gas cylinder level indicator deactivated
4.6.2
Settings
1. Select
> Station gas configuration > Settings.
Fresh air inlet setting
Expert settings
Gas cylinder level indicator
To set the fresh air input:
1. Select Fresh air.
2. Select desired setting:
Supplied by pump - fresh air inlet (default setting)
Supplied by compressed air input - compressed air
inlet
3. Confirm selection with OK.
Dräger X-dock 5300/6300/6600
Gas pressure too low.
Gas pressure correct.
Gas pressure too high.
Gas cylinder level
61 - 100 %
30 - 60 %
<30 %
►
►
►
The following settings can be made in the expert menu:
Ignore max. concentration for quick bump test
Set test behaviour for missing test gases
To ignore the concentration limit specified by Dräger for the
quick bump test:
1. Select Ignore max. conc. for BTQ.
2. Activate checkbox (default setting: deactivated).
3. Confirm selection with OK.
Activating this feature allows users to use higher test gas
concentration for the quick bump test than recommended by
Dräger.
WARNING
!
This feature must only be activated by trained and
qualified personnel since wrongly selected test gas
concentrations may lead to a positive test result
although the gas detector's warning comes too late.
To set the test behaviour with missing test gases:
1. Select Cancel the test on missing gas.
2. Activate checkbox (default setting: activated).
3. Confirm selection with OK.
This function can be used to set whether or not a test or
calibration is carried out when a required test gas is not
connected.
WARNING
!
If this function is deactivated, the corresponding
channel is not tested or calibrated.
To set the gas cylinder level indicator:
NOTICE
i
i
The gas cylinder level indicator is only available for
cylinders that are configured via a Dräger part no.
1. Select Gas level monit.
2. Activate or deactivate check box Gas level monit.
3. Confirm selection with OK.
To reset the gas cylinder level indicator for a new test gas
cylinder:
1. Connect a new test gas cylinder to a test gas connection.
2. Select
> Station gas configuration.
3. Select desired gas inlet.
4. Select Further options and select
cylinder level indicator.
Basics
to reset the gas